Veo Mobile Guia do Utilizador Página 205

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 262
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 204
Alarm conditions
Alarms put the pump in Attention mode.
A (ALARM)
This alarm shows the letter A followed by two numbers. A-alarms cause all insulin delivery to stop. Your
pump settings are retained. If this alarm repeats often, call your local Medtronic Diabetes representative for
assistance.
AUTO OFF
Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature,
and so insulin delivery has been stopped.
BATT OUT LIMIT
Occurs if the battery has been out of the pump for more than five minutes. Verify that the pump time and
date are correct. If not correct, go to the UTILITIES MENU and reset the date and time.
BOLUS STOPPED
If this alarm occurs, it is very important to check bolus history to review how much of the bolus was actually
delivered. If necessary, reprogram a bolus with the amount not yet delivered.
BUTTON ERROR
Occurs if a button has been continually pressed for more than three minutes.
Troubleshooting, alerts and alarms 185
Chapter 13
Vista de página 204
1 2 ... 200 201 202 203 204 205 206 207 208 209 210 ... 261 262

Comentários a estes Manuais

Sem comentários