Veo Mobile Guia do Utilizador Página 82

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 262
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 81
Filling the reservoir
1 Remove reservoir from package. Make sure plunger rod is fully extended.
2 Swab vial with alcohol.
3 Making sure you do not push down on the plunger, press the transfer guard onto the vial.
plunger
rod
transfer
guard
reservoir
vial
31
4 Push down on the plunger to pressurize the vial.
5 While still holding the plunger rod, flip the vial over so the vial is up, slowly pull down on the plunger
to fill the reservoir.
6 Gently tap the side of the reservoir to make any air bubbles rise to the top of the reservoir.
45
6
7 Slowly push up on the plunger just enough to remove any air bubbles from the reservoir.
8 Slowly pull down on the plunger to fill the reservoir to the number of units desired.
9 With the vial down, hold the transfer guard. Turn the reservoir counter-clockwise, then pull straight up
to remove the vial from the transfer guard.
Starting on insulin62
Vista de página 81
1 2 ... 77 78 79 80 81 82 83 84 85 86 87 ... 261 262

Comentários a estes Manuais

Sem comentários