Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 95
7.3 Warning and Error Codes
Each time a fault or error occurs during the operation of the detector, a warning code or
error code appears on the detector display. In this case, the Continue soft key appears on
the display for navigation
If you operate the detector from a chromatography data system, a message will also be
documented in the audit trail of the software.
Warning Codes
The following table lists the detector-related warning codes and messages along with
recommended remedial actions. Warning codes will neither interrupt data acquisition from the
detector nor interfere with automated sequences or batch runs when you operate the detector
from a chromatography data system.
Note the warning code or the exact wording of the message and contact the Thermo Fisher
Scientific Service for Dionex HPLC products if you are unable to eliminate the problem.
Warning Code Warning Message Description and Remedial Action
15000 Charger voltage
failed
The voltage of the charger approaches the maximum level.
Turn the detector off and on. If the voltage still fails, contact
Service.
15010 Leak sensor fault
detected
The leak sensor has reported a leak.
Check the nebulizer and/or stream-switching valve for signs
of leakage. Dry the leak sensor (→ section 8.4, page 112).
15008 Stream-switching
valve (SSV) failure
Corona Veo RS only
The SSV has failed to rotate. From the System Setup Menu,
manually actuate the valve. If the valve still fails to rotate,
disassemble and inspect the SSV (→ section 8.7.1,
page 116). If the error persists, replace the rotor and stator
(→ section 8.7 , page 116).
15011 High engine
temperature
The internal temperature of the detector is too high.
Turn the detector off and on. Test the evaporation heater by
the self test. Allow the detector to come to thermal
stabilization. If the heater still fails, contact Service.
15012 Low engine
temperature
The internal temperature of the detector is too low.
Turn the detector off and on. Test the evaporation heater by
the self test. Allow the detector to come to thermal
stabilization. If the heater still fails, contact Service.
15013 High evaporation
temperature
The temperature of the evaporation tube is too high. Set a
lower temperature in the specified range (→ section 5.6.3,
page 66).