Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 121
9 Technical Information
Specification Corona Veo Corona Veo RS
Detection Type
Charged Aerosol Detection
Nebulization
FocusJet™ concentric flow design
Mobile Phase Flow Rate
0.2 2.0 mL/min 0.012.0 mL/min
Filter Time Constants
0.1, 0.2, 0.5, 1.0, 2.0, 3.6, 5.0, 10.0 seconds
Evaporation Temperature
Selectable values:
35 °C or 50 °C
Settable range:
ambient +5 °C to +100 °C
Temperature Accuracy
< ± 5 °C
Equilibration Time
< 30 minutes to 35.0 °C evaporation temperature
Integrated Stream-Switching
valve
N. a. 6-port 2-position valve
(controllable via TTL input)
Digital Noise Filtering
4
th
order low-pass Bessel filter
Gas Requirements
Compressed air or nitrogen
Gas Pressure Control
Manual Electronical
Gas Supply Pressure
480 550 kPa (7080 psig)
Data Collection Rate
max. 100 Hz (Chromeleon 6.80) max. 100 Hz (Chromeleon 6.80) or
200 Hz (Chromeleon 7)
User Input/Display
Integrated liquid crystal color display with touch screen with menu
functions
Software Control
Available detector functions are software controllable via USB
(USB 1.1 or 2.0 compatible).
Analog Signal Output
(optional)
BNC analog signal output
Analog Output
0 – 1 V DC
Full-Scale Analog Output
Range
1 pA to 500 pA in 1-2-5 sequence
Rear Interface
AC connector, power switch, USB port, Digital I/O terminals (4 TTL
inputs, 3 relay outputs)
I/O Interfaces
4 TTL inputs (Gas Off, Autozero, Start, Stream-Switching Valve), 3
relay outputs (one fixed labeled 'Pump Off')
Safety features
Leak sensor, liquid sensor, gas pressure relief valve
Power requirements
100-240 V AC, 50/60 Hz, 100 VA