Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 130 Operating Instructions
Important: When you connect the TTL inputs, ensure that the potential across
the pins is set to 5 V DC ±0.25 V. When you connect relays outputs,
the potential across the pins must be less than 30 V DC and the
current must be less than 0.5 A.
Important: Lors de la connexion des entrées TTL, assurez-vous que le potentiel
aux broches est défini à 5 V CC ±0,25 V. Lors de la connexion des
sorties, le potentiel aux broches doit être inférieure à 30 V CC et le
courant doit être inférieure à 0,5 A.
TTL Inputs
The inputs are TTL compatible and are assigned "Gas Off", "Autozero", "Start" and
"Divert". Use a true contact closure or relay output for the inputs on the Digital I/O
terminals. If a TTL level change is used, the voltage must be between 0 and 5 V, and the
polarity sense of the device must match that of the detector (i.e. the "+" and the "-" of the
external device must be connected to the "+" and the "-" of the detector, respectively). The
"+" port is the top one in each group and the "-" port is immediately below it, followed by
ground (GND).
• Start
The "Start" input is available for use with a chromatography data system that requires
triggering event synchronization for data collection from an external device, such as an
autosampler. The detector requires a contact closure or shorting of the terminals (a signal
of at least 0.5 seconds) from the external device.
• Gas Off
The "Gas Off" input can be used to turn off the internal gas inlet valve and stop the gas
supply to the nebulizer. It is recommended to use this input only when the gas supply is
limited.
If the" Gas Off" input is activated, the detector will shut down internal gas supply. This
will result in an alarm condition. Stop the alarm condition using the detector display or
the chromatography data system.
• Autozero
The "Autozero" input can be used to set the signal of a channel to a "zero point"
regardless of the current that is being measured.
Allow sufficient time after you performed the Autozero and before injection to ensure
that the autozero command is completed. Alternatively, an autozero may be performed
during a run. In this case, make sure that autozero occurs at a point in the chromatogram
where the response curve is expected to be reasonably flat (i.e., 15 seconds prior to the
elution of a peak).
• Divert
The "Divert" input can be used to control the stream-switching valve with a third-party
driver.