Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 28 Operating Instructions
Systems with Series and Parallel Detector Configuration
The detector can be used in series or parallel configuration with other detectors (for example,
a UV detector or mass spectrometer).
Series Configuration
In series configuration, one flow path from the pump is used for each type of detector
installed on the system. Consider the following:
In general, charged-aerosol detectors such as the Corona Veo detector must be placed last
in the flow path within the system.
The addition of other detectors may contribute to band broadening, and have an adverse
effect on peak shape measured by the detector.
Parallel Configuration
In parallel configuration with another detector or a mass spectrometer, consider the following:
Use a flow splitter to ensure balanced distribution of liquid flows to each detector in the
system.
Tip: Thermo Fisher Scientific recommends using an adjustable analytical flow
splitter (part no. 70-6377, adjustable from 1:1 to 20:1) for flexible,
accurate and precise liquid flow.
When you use a passive flow-splitting device and thus adjust (balance) the liquid flows,
consider the backpressure from the detector. Observe the flow rate specifications in the
'Technical Information' section ( section 9, page 121).