Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 102 Operating Instructions
Problem Probable Cause Remedial Action
High noise, non-periodic
baseline fluctuation
(Cont'd)
There are temperature fluctuations
in the laboratory.
Check the thermostat in the
laboratory. If possible, note the
temperature fluctuations/values in
the system log.
Periodic baseline
fluctuation, pulsation
There are pressure fluctuations
from the pump.
Purge the pump; check general
function ( Pump Manual).
There are air bubbles in the
system.
Purge the system
( Pump Manual).
Peak Tailing Too large extra column volume Use short capillary connections with
a suitable inner diameter.
There are bad capillary
connections.
Use different capillaries, for
example, Viper capillaries.
Peak Broadening,
increased dead time
The inner diameter of the
capillary to the detector is too
large.
Change the capillary.
The filter frits on the solvent lines
are clogged.
Check the filter for permeability.
Replace the filter frit if necessary
( Pump Manual).
The capillaries are clogged or
capillary connections bad.
Replace the capillaries. Use
different capillaries, for example,
Viper capillaries.
The sample loop is clogged.
Replace the needle ( Autosampler
Manual).
The proportioning valve is
defective.
Contact Service.
The column is overloaded or
contaminated.
Clean or replace the column.
The eluent has changed. Use fresh solvent.
Reproducible ghost
peaks in the
chromatogram.
The degassing channels are
contaminated.
Rinse the degassing channels
(
Solvent Rack or Pump Manual).
The solvents are degraded or dirty
or their purity is insufficient.
Use fresh and appropriate solvents.
Contamination occurs in the
autosampler or somewhere in the
system.
Flush the autosampler or the entire
system using an appropriate solvent.
Some broad ghost peaks
in the chromatogram.
Late eluting peak from previous
analysis.
Extend the run time. Increase the
elution strength of the gradient
(higher organic content). At the end
of the run, flush column with strong
eluent.