Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 34 Operating Instructions
4.2 Connecting the Drainage
The detector has two drain ports at the bottom right of the enclosure to collect liquid leaks
or waste from the detector. Use the drain tubing shipped with the detector to direct liquid
from the two drain ports to waste.
Fig. 5: Drain ports
No. Description
1 Leak port
For liquid leaks from the drip tray.
2 Waste port
For liquid from the drain pump during operation.
When connecting the drain ports of the detector, observe the following:
• Connect the ports separately with a drain line for each port. Do not use a connection tube
to connect both ports with each other.
• Both drain ports may be directed to the same waste.
• Use the elbow and tee pieces from the accessories kit for the detector to connect the drain
ports.
• Make sure that the tubing and waste bottle are below the height of the detector.
Drainage in an UltiMate 3000 system
In an UltiMate 3000 system, you can use the components from the drain kit for
UltiMate 3000 systems to direct liquid leaks to waste via the drain system. The kit is
shipped with the UltiMate 3000 pumps and can be ordered separately (part no. 6040.0005).
The kit includes all required components and detailed installation instructions.
If there is more than one UltiMate 3000 detector in your system and you need an additional
tee piece, you can find one in the accessories kit of the UltiMate 3000 fluorescence,
multiple wavelength, or diode array detector.
2
1