Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 80 Operating Instructions
6.1.2 Installing the Detector
The detector should complete the self-test prior to connecting the detector to the computer
to prevent possible installation errors. When the display shows the Main Menu, the self-
test is complete, and that the detector is ready for connection.
After the USB software for the detector has been installed ( page 78), install and
configure the detector in Chromeleon:
1. Start the Chromeleon Server Monitor ( page 78) and the Chromeleon server if they
are not yet running.
2. Start the Chromeleon Server Configuration program by clicking Start on the taskbar.
Point to Programs (or All Programs, depending on the operating system), point to
Chromeleon, and then click Server Configuration.
3. If necessary, click the plus sign beside the server icon to display the items
underneath.
4. Select the timebase to which the detector will be assigned, or create a new timebase
(on the Edit menu, click Add Timebase).
5. Open the Add device to timebase dialog box. To do so, click Add Device on the Edit
menu or right-click the timebase and click Add Device on the menu.
6. On the Manufacturers list, click Dionex and on the Devices list, click Corona Veo
Charged Aerosol Detector.
7. The configuration pages are opened. On each page, verify that the settings are correct
and select additional settings if needed. For a description of the pages, see
section 6.1.3.1 ( page 81).
8. Click OK to complete the configuration of the detector.
9. On the File menu, click Save Installation and then close the Server Configuration
program.