Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 45
5 Operation
5.1 Safety Guidelines for Operation
English Safety Guidelines
For the safety guidelines in French, see page 46.
Important: When operating the detector, observe the following safety
guidelines:
• Always turn on the gas supply before starting the flow of mobile phase.
• When you stop the gas flow, always turn off the mobile phase flow and allow the
detector to remain pressurized for 5 minutes before your turn off the gas flow.
• Turn on the gas flow only when a nebulizer is installed.
• Observe the pressure range for the gas supply (→ section 9, page 121) and the internal
gas pressure setting as stated in the Installation Instructions for the nebulizer.
• Operate the detector only when it is connected to appropriate ventilation. Exhaust gas
ventilation should be at atmospheric pressure.
• If a leak occurs, turn off the detector immediately, stop the pump flow, and remedy the
situation.
• If the pressure relief valve is open and leaking gas, immediately shut down the detector.
For details, see section 5.7.3 (→ page 71).
• After operation of the detector
If you operated the detector with organic solvents in high concentration that may form
peroxides upon evaporation (un-stabilized THF, 2-propanol, etc.), flush the flow paths
completely with HPLC-grade water prior to shut down.