Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 119
8.9 Updating the Detector Firmware
The detector is shipped with the most recent firmware version. The firmware version is
displayed on the Main Menu screen.
To update the detector firmware, the Chromeleon Chromatography Management System is
required. The following update description refers to the Chromeleon 6.80 Service
Release 13.
To check which firmware version is installed in the detector and which version is included
in Chromeleon:
In the Chromeleon Server Configuration program, open the configuration pages for the
detector ( page 81). On the General page, the firmware version is displayed.
In the Windows Explorer, locate the IQReport.log file in the IQ folder of your
Chromeleon installation. In the file, search for CAD3 FW (.run file).
Tip: When you update the firmware via Chromeleon, this information will also be
provided during the download ( below).
Whenever a new firmware version is released for the detector, the new version will be
provided with the next Chromeleon Service Release and described in the related release
notes.
The new firmware will not be downloaded automatically to the detector when you install a
Chromeleon Service Release. To update the detector firmware, follow these steps:
Important: To ensure that the download is successful, make sure that the
communication between the detector and Chromeleon is not
interrupted during the download, and do not turn off the detector.
Important: Au cours du téléchargement, assurez-vous que la communication
entre la pompe et Chromeleon n'est pas interrompue et n'arrêtez pas
l'instrument. Ceci peut entraîner des dysfonctionnements de
l'instrument.
1. Before you begin, verify that
The detector is connected in Chromeleon.
The Chromeleon server is in running idle mode. This means that all processes on the
Chromeleon server PC and in Chromeleon have been stopped.
2. Start the Server Configuration program ( page 80).
3. Right-click the detector in the timebase and select Properties on the menu.