Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 98 Operating Instructions
Error Code Error Message Description and Remedial Action
29003 High flow ratio The gas flow ratio is outside of specification for more than 60
seconds.
The contact closure was sent to the Digital I/O terminal of the
detector. If a pump is connected to the Digital I/O, the pump flow
will be stopped. On the detector display, press any soft key to
continue.
Corona Veo RS only: The stream-switching valve will divert liquid
flow away from the nebulizer to prevent detector flooding.
Check that
• gas supply pressure is in the range of 480 kPa – 550 kPa
(70 – 80 psig) and that gas is still being delivered to the
detector.
• gas pressure reading on the detector display is the range of
379 kPa – 448 kPa (55 – 65 psig).
Turn the detector off and on and allow the self test to test the
pressure sensors.
If the error persists, then there may be a possible gas blockage
within the detector. Contact Service.
29009 Liquid sensor
fault detected
Liquid is present the mixing chamber of the detector (→ Fig. 29).
The contact closure was sent to the Digital I/O terminal of the
detector. If a pump is connected to the Digital I/O, the pump flow
will be stopped. On the detector display, press any soft key to
continue.
Corona Veo RS only: The stream-switching valve will divert liquid
flow away from the nebulizer to prevent detector flooding.
Turn the detector off and on and allow the self test to test the
pressure sensors.
In the Main Menu, set the gas pressure to On and attempt to dry
the liquid present inside the detector for 24 hours. If the error
persists, contact Service.
29000 Low ion trap
voltage
The ion trap voltage is too low. Liquid may be present in the ion
trap. The voltage may have been shorted due the detector flooding.
The contact closure was sent to the Digital I/O terminal of the
detector. If a pump is connected to the Digital I/O, the pump flow
will be stopped. On the detector display, press any soft key to
continue.
Corona Veo RS only: The stream-switching valve will divert liquid
flow away from the nebulizer to prevent detector flooding.
Turn the detector off and on and allow the self test to test the
pressure sensors.
In the Main Menu, set the gas pressure to On and attempt to dry
the liquid present inside the detector for 24 hours. If the error
persists, contact Service.
For more information about the internal safety features of the detector, see section 5.7
(→ page 70).