Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 63
The following table is to assist you with typical flow rates for certain column diameters:
Column Diameter Typical Flow Rate
4.6 mm 1.0 mL/min
3.0 mm 0.5 mL/min
2.0 mm 0.2 mL/min
1.0 mm 0.05 mL/min
Fig. 16: Typical flow rates for various columns
5. Allow the detector to equilibrate until you observe a steady baseline.
Monitor the background signal of the detector. It will increase initially and then start to
fall. Over a short period of time the signal will fall to a low level and then stabilize.
6. The detector is ready for data acquisition when the baseline is stable. The time
required to obtain a stable baseline depends on the application.
7. Autozero the detector: From the Run Mode Menu, press the Autozero soft key.
8. The detector is now ready for sample analysis.