Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 131
Relay Outputs
The "Pump Off" output contact is used to send a signal to the pump (or other external
device) to stop the flow of mobile phase and turn off the pump. This signal may be used to
turn off the pump to prevent detector from being flooded with mobile phase if the gas
pressure falls below 50 % of the required gas volume for more than 60 seconds.
To use the "Pump Off" contact closure, the HPLC system must be able to accept an input
from the detector that can be programmed to turn off liquid flow. In some cases, a special
cable is required.
The pins labeled "CC1"' and "CC2" provide contact closures that can be used to start an
external device (for example, an autosampler) or to change a valve position. The two ports
act in a similar fashion to the "Pump Off" output. When the output contact closures are
used, connect the outputs according to their polarity ("+" or "-"). Connect the third wire on
the Digital I/O cable (part no. 70-4850) to the ground of the pin.