Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 27
3.3 Positioning the Detector in an UltiMate 3000 System
If the detector is part of an UltiMate 3000 system, for example for analytical HPLC
applications, stack the individual modules and interconnect them on the rear panel as
shown in the example in Fig. 4.
Solvent
rack
(optional)
Pump
Auto-
sampler
Column
compart-
ment
Detector
Fig. 4: Example for stack arrangement and rear panel connections
in an UltiMate 3000 system with the Corona Veo RS detector
Observe the following when positioning the detector:
• The arrangement of the system modules depends on the application and may vary if an
optical detector is used additionally.
• Place the detector in a position so that the connections between the output of the column
and the inlet of the detector are minimized. This will reduce post-column band
broadening effects and optimize chromatographic resolution.
• In an UltiMate 30000 system, apart from the Solvent Rack, all system modules can be
connected separately to the computer via the USB port on the rear panel of the module,
respectively. However, Thermo Fisher Scientific recommends interconnecting all
modules, and then connecting the system to the computer with only one connection.