Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 53
5.3.3.2 Graph Screen
The Graph screen displays the online detector signal during data acquisition.
Fig. 10: Graph screen (here: Corona Veo RS)
The following functions and parameters are available on the screen:
Soft key Select to…
Gain ± Change the default signal range. This function refers to the online plot range as is
displayed.
Rate ± Update the display rate. Rate is displayed in seconds per division. A rate of 4.7
indicates each horizontal division is 4.7 seconds and the entire chromatogram is 47
seconds. With this function you can change the default time constant window of the
online plot.
Auto Adjust the auto-rate/gain setting. The online plot will scale the signal automatically
to the peak with the largest magnitude on the online plot.
Exit Exit out of the Graph screen.
GRAPH DISPLAY
Exit
Gain+
Run Mode
Diags
Evap T
Current = 0.000 pA Stat = Normal
Gain = 16 Rate = 3.1
Gain-
Rate+
Rate-
Auto