Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 60 Operating Instructions
2. If applicable
In the Run Mode Menu, select the Exit soft key to return to the Main Menu.
Tip: If you operate the detector from a chromatography data system, you may
need to disconnect the detector in the system in order to access the Main
Menu.
3. In the Main Menu, select the System Menu soft key.
4. Select the Gas soft key to display the current gas pressure value.
5. Before adjusting the pressure, loosen the lock nut of the gas regulator knob on the rear
panel of the detector.
6. Adjust the gas pressure:
♦ To increase the gas pressure, turn the regulator knob slowly clockwise.
♦ To decrease the gas pressure, turn the regulator knob slowly counterclockwise.
7. While turning the knob, monitor the increase or decrease of the gas pressure on the
detector display.
8. Re-tighten the lock nut of the regulator knob. Ensure that the indicated pressure
reading on the display does not change. Hold the regulator knob while tightening the
lock nut. This will ensure that the pressure does not change while you tighten the lock
nut.
9. Wait until the gas pressure has stabilized and reached the set point. Monitor the gas
pressure (→ see further down in this section).
Setting the Gas Pressure for the Corona Veo RS
Regulate the internal gas pressure from the display to provide a consistent nebulizer
pressure for a stable aerosol.
1. If applicable
Stop the mobile phase flow from the pump and wait several minutes for the detector
pressure to stabilize.
2. If applicable
In the Run Mode Menu, select the Exit soft key to return to the Main Menu.
Tip: If you operate the detector from a chromatography data system, you may
need to disconnect the detector in the system in order to access the Main
Menu.
3. In the Main Menu, select the System Menu soft key.
4. Select the Gas soft key to display the current gas pressure value.