Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 67
Factor
Evap. Temp. Guidelines
Effect
Corona Veo Corona Veo RS
Mobile phase
composition
volatile modifiers (e.g.
TFA, formic acid,
ammonium formate)
35 °C or
50 °C
non-volatile modifiers
(e.g. sodium phosphate)
Do not use ---
Mobile phase
flow rate
35 °C minimal
Analyte
Non-volatile 50 °C 35 °C or 50 °C minimal
Semi-volatile 35 °C
Use lower or
higher temperature
settings
Detector
response,
baseline noise
5.6.4 Using Gas Flow Modes (Corona Veo RS only)
The Corona Veo RS detector is capable of two separate gas pressure modes: analytical gas
flow mode and microLC gas flow mode.
Each of these modes sets a specific pressure, based upon the default gas pressure of the
nebulizer. They are designed as a simple way to optimize the nebulizer efficiency when
you use very high (analytical) or very low (microLC) flow rates that range from
0.01 mL/min – 2.00 mL/min. These modes of operation are only available when you use
compatible chromatography software.
Analytical gas flow mode:
The gas pressure set point corresponds to the default gas pressure setting as described in
the certificate that is shipped with the nebulizer.
MicroLC gas flow mode:
The gas pressure value is calculated by multiplying the analytical gas flow mode value by
a factor in order to optimize the gas pressure for certain types of liquid flow rates and
mobile phase compositions.