Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 36 Operating Instructions
4.4 Making Flow Connections
The following section provides information about how to connect capillaries to your
detector.
4.4.1 General Guidelines
When connecting capillaries to the detector, observe the following general precautions:
• Observe the precautionary statements for capillaries and capillary connections in
section 1.2.2 (→ page 4).
• When you connect capillaries, make sure that the connectors are free from contaminants.
Even minute particles may cause damage to the system.
• Use only the capillaries shipped with the detector or original spare capillaries.
• Always make sure that the ID of the replacement capillary corresponds to the ID of the
capillary shipped with the system.
• Make sure that the connectors and capillary ends are free from contaminants and not
damaged. Even minute particles may cause damage to the system, for example, to the
nebulizer and, with the Corona Veo RS, the stream-switching valve.
• The connection between the column and the detector should be as short as possible to
prevent band-broadening.
For more information about the available capillaries, see section 10.2 (→ page 126).