Veo Observer Plus Instruções de Operação Página 81

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 80
Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 73
Corona Veo RS: The stream-switching valve will actuate and divert the liquid flow away
from the detector.
The drain pump will turn on automatically.
Tip: When you operate the detector from a chromatography data system, an error
message will also be documented in the audit trail of the software.
To remedy the problem:
Check the gas supply for leaks (if appropriate) and make sure that the gas supply is
properly connected.
Set the gas pressure to the setting specified for your nebulizer ( page 59).
Reset the pump and re-start the analysis.
Vista de página 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 143 144

Comentários a estes Manuais

Sem comentários