Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 114 Operating Instructions
3. Carefully move the cover plate away from the detector enclosure and place it close to
the detector.
4. With the multi-tool from the accessories kit for the detector, push in the locking ring
on the push-in fitting and pull carefully on the gas tubing.
5. Carefully pull the filters from out of the clamps on the cover plate.
6. To install the new filters, follow the above steps in reverse order. Install the gas filters
in the direction of the gas flow through the filters. Flow direction signs are indicated
on the filter housings.
7. Re-connect the power cord to the main power receptacle.
8. Turn on the detector. Observe the notes and precautions in section 5.2 ( page 47).
Allow the detector to perform the self test before proceeding.
9. On the detector display, reset the gas filter replacement date from the System Setup
Menu.
a) Select Date to get to the Date & Time Setup Screen.
b) Select Reset Filter Chg Date. The replacement interval is automatically set for 12
months after replacement and can be reset manually.