Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Operating Instructions Page 33
4 Preparation for Operation (Startup)
4.1 Overview of Actions
After you have unpacked, positioned and connected the detector as described in
sections 3.1 through 3.4 (→ page 23 and following), prepare the detector for operation.
Follow the sequence of steps below:
1. Connect the drain tubing as described in section 4.2 (→ page 34).
2. Connect the gas tubings to the gas inlet and the gas exhaust as described in section 4.3
(→ page 35).
3. Install the nebulizer as described in the Installation Instructions for the nebulizer.
4. Make the flow connections from the detector to the other modules of your system, as
required by your application. For information about the flow connections, see
section 4.4 (→ page 36).
5. Turn on the detector. Observe the notes and precautions in section 5.2 (→ page 47).
Allow the detector to perform the self test before proceeding.
6. Chromeleon installation (optional)
When the display shows the Main Menu after the self test, set up the detector in
Chromeleon as described in section 6.1 (→ page 78).
7. If required
Adjust the brightness of the detector display (→ section 5.3.2, page 49).
8. If required
Set date and time on the detector display (→ section 5.3.4.2, page 57).
9. Set the gas pressure for the nebulizer as described in section 5.4.1 (→ page 59).
10. Before using the detector for sample analysis, prepare the LC system and the detector
as described in sections 5.4.2 and 5.4.3 (→ starting from page 62).